FINAL REPORT SIM COMPARISON IN MASS STANDARDS
SIM.M.M-K5
Kornblit, F1; Leiblich, J.1 ; Becerra, L.O.2; Peña, L.M.2; Luján,L.2; Díaz,J.C.2; Centeno, L.M.2; Ramos, O3; Rodríguez, S.3; Santo, C.4; Caceres,
J.4; García, F.5; Leyton, F.5; Claude J.6; Loayza, V.M.7; Cacais,F.A.7
Abstract This report summarizes the results of a SIM comparison of mass standards carried out between 7 NMIS. Five standards with nominal values 2 kg, 200 g, 50 g, 1 g and 200 mg have been circulated by the NMIs. The results reported by the participants are consistent with each other and with the key comparison reference value of comparison CCM.M-K5 to which this comparison has been linked.
1. General Information The present comparison, named SIM.M.M-K5, was planned and carried out in order to evaluate the degree of equivalence in the calibration of high accuracy mass standards, and to provide evidence supporting CMCs claimed by the participants in high accuracy mass calibrations delivered by them. It is part of a more general project which comprises three comparisons:
SIM.M.M-K1 for mass calibration of nominal value 1 kg SIM.M.M-K5 for mass calibration of nominal values 2 kg, 200 g, 50 g, 1 g and 200 mg SIM.M.D-K3 for volume determination of stainless steel weights of 2 kg, 1 kg, 200 g,1 g
2. Data of the participant NMIs and Technical Contacts The following SIM institutes have participated in the comparison:
Institute LACOMET LATU INTI CESMEC CENAM NRC-INMS INMETRO
Country Costa Rica Uruguay Argentina Chile México Canada Brazil
Technical Contact(s) Ramos, O; Rodríguez, S. Santo, C.; Cáceres, J. Kornblit, F; Leiblich, J. García, F.; Leyton, F. Becerra, L.O.; Peña, L.M.; Luján, L.; Díaz,J.C.; Centeno, L.M. Jacques, C. Loayza, V.M.; Cacais, F.A.
INTI (Argentina)1 has acted as the pilot laboratory
1INTI, Instituto Nacional de Tecnología Industrial (Argentina), contact e-mail: ferk@inti.gob.ar, 2 CENAM, Centro Nacional de Metrología (México); 3LACOMET, Laboratorio Costarricense de Metrología; 4 LATU, Laboratorio Tecnológico del Uruguay (Uruguay); 5CESMEC (Chile); 6NRC, National Research Council (Canada), 7 INMETRO, Instituto Nacional de Metrologia, Qualidade e Tecnologia (Brazil)
1/13
3. General Considerations and Procedure A set of five stainless steel standards, made by Masstech and provided by CENAM was used for the comparison. The nominal values and identifications are shown in Table 1. In June 2012, once measurements were finished, CENAM determined the volume of the standards referred to 20 ºC, V. These results and the corresponding standard uncertainties uV are shown in Table 1.
Table 1. Data associated to the standard weights
Nominal value
Serial Number
Identification
V / cm3 uV / cm3
2 kg
1867 540161558204 251,051 0,024
200 g
1865 540161556204 25,1487 0,0024
50 g
1864 540161555204 6,2624 0,0008
1g
1863 540161554204 0,1258 0,0005
200 mg
S/N 540140212102 0,025 0 0,0002
The traveling standards were placed in individual wooden cases for transportation purposes, which were placed in a carrying transportation case, jointly with the standards corresponding to the comparisonsSIM.M.M-K1 and SIM.M.D-K3. In all the cases, the transportation among laboratories was made by hand, by technical staff of the NMIs.
A protocol was agreed previously to the comparison. In it, instructions to travel, initial inspection in each country, store, handling and acclimatization of the standards have been specified. Particularly, the following criteria were agreed:
• The standards were not washed during the comparison.
• The institutes applied their own method to measure the mass of the standards, in order to achieve uncertainties as low as possible, according to its capabilities. That means that each laboratory applied subdivision methods with different weighing designs.
• The CIPM-2007 formula [1] was applied by all the participants to determine the air density, in order to estimate buoyancy effects.The buoyancy corrections were applied by the participants retrospectively after the circulation of the transfer standard.
4. Schedule The measurements followed the schedule shown in Table 2.
2/13
Table 2. Measurement schedule
N° Institute / Country
Dates of the measurements
-- CENAM / Mexico
July 2009
1 LACOMET / Costa Rica October 2009
2 LATU / Uruguay
January 2010
3 INTI / Argentina
April 2010
4 CESMEC / Chile
July 2010
-- CENAM / Mexico
June 2011
5 NRC /Canada
February 2011
6 INMETRO/Brazil
January 2012
7 CENAM / Mexico
July 2012
5. Stability of the standards The drift of the standards of 2 kg, 200 g, 50 g, and 1 g was evaluated from three series of measurements performed by CENAM, in July 2009, in June 2011, and in July 2012. The standard of 200 mg was only measured in June 2011 and July 2012.The mass errors e obtained their associated uncertainties and normalized errors values En are listed in Table 3 are plotted in Figures 1A to 1E (the uncertainty bars in them correspond to k = 2).
Table 3. Stability of the standard. Values obtained by CENAM
Date
e / mg
U (k=2)
En
July 2009
2,79
0,17
-0,1
2 kg
June 2011
2,83
0,13
0,1
July 2012
2,82
0,13
0,0
July 2009
0,305
0,014
-0,4
200 g
June 2011
0,324
0,016
0,3
July 2012
0,317
0,014
0,1
July 2009
0,076
0,004
0,0
50 g
June 2011
0,076
0,006
0,0
July 2012
0,077
0,005
0,1
July 2009
0,004
0,001
0,0
1g
June 2011
0,004
0,001
0,0
July 2012
0,004
0,001
0,0
200 mg
June 2011 July 2012
0,000 1 -0,000 2
0,000 7
0,2
0,000 7
-0,2
It can be concluded that no significant effects associated to drifts of the standard need to be considered.
3/13
6. Summary of the reported results The results sent by the participants are expressed as the mass error e of each standard, taken from their nominal values. They are shown in Table 4, as well as the associated uncertainties U (for k = 2).
Table 4. Mass errors e as reported by the participants, and the corresponding uncertainties U
2 kg
200g
50 g
1g
200 mg
e/mg U/mg e/mg U/mg e/mg U/mg e/ µg U/ µg e/ µg U/ µg
LACOMET 2,54 0,84 0,312 0,020 0,080 0,022 3,2
3,0 -0,2 1,3
LATU
2,52 0,68 0,315 0,024 0,071 0,006 4,4
2,6
0,4
1,6
INTI
2,87 0,20 0,315 0,018 0,077 0,010 4,3
2,0 -0,4 1,2
CESMEC 2,50 0,30 0,315 0,030 0,074 0,010 4,3 3,0 0,0 2,0
CENAM
2,82 0,13 0,317 0,014 0,077 0,005 4,2
1,0
0,0
0,7
NRC
2,66 0,11 0,317 0,028 0,073 0,014 3,4 1,4 -0,3 1,4
INMETRO 2,77 0,22 0,312 0,022 0,070 0,006 3,8 1,4 -0,2 0,6
NLRV
2,73 0,07 0,315 0,008 0,073 0,0029 3,9 0,6 -0,1 0,4
7. Data consistency and computation of (non-linked) reference values In order to check the consistency of the results, 2 tests as proposed in [2] were applied. The
assumptions to employ the so-called Procedure A were considered valid. For the different
standards, the sums of squares corresponding to the whole set of results yielded the following values of the 2 statistic:
Nominal value
2 obs
2 kg
8,6
200 g
0,3
50 g
4,6
1g
1,4
200 mg
1,0
while the corresponding critical value for 6 degrees of freedom and significance level = 0,1 is 10,6. So, the data were considered consistent in all the cases. Following [2] the weighted averages were calculated and established as the (non-linked) reference values of the comparison, called NLRV. They are shown in the last row of table 3, as well as their associated uncertainties.
Then, for each participant and standard, degrees of equivalence D and normalized errors En were calculated according to 1:
D e NLRV ;
UD 2
u2 u2 ; NLRV
En D U D
(1)
4/13
being e,u the results reported by the participant. D, UD and En are shown in Tables 5A and 5B. The values of D and their associated uncertainties are plotted in figures 2A to 2E. In each plot, Here, NLRV is represented as the zero line and their associated uncertainty as solid pink lines.
Table 5A. Degrees of equivalence respect to the non-linked reference value
2 kg
200 g
50 g
1g
200 mg
D / mg UD / mg D / mg UD / mg D / mg UD / mg D / µg UD / µg D / µg UD / µg
LACOMET -0,19 0,84 -0,003 0,018 0,007 0,022 -0,7 2,9 -0,1 1,2
LATU
-0,21 0,68 0,000 0,023 -0,002 0,006 0,5 2,5 0,5 1,5
INTI
0,14 0,19 0,000 0,016 0,004 0,010 0,4 1,9 -0,3 1,1
CESMEC -0,23 0,29 0,000 0,029 0,001 0,010 0,4 2,9 0,1 2,0
CENAM 0,09 0,11 0,002 0,012 0,003 0,005 0,2 0,8 0,1 0,6
NRC
-0,07 0,08 0,002 0,027 0,000 0,014 -0,5 1,3 -0,2 1,4
INMETRO 0,04 0,21 -0,003 0,021 -0,004 0,005 -0,1 1,3 -0,1 0,5
Table 5B.Values of En respect to the non-linked reference value
2 kg 200 g 50 g 1 g 200 mg
LACOMET -0,2 -0,2 0,3 -0,2 -0,1
LATU
-0,3
0,0
-0,4
0,2
0,4
INTI
0,8
0,0
0,4
0,2
-0,2
CESMEC -0,8
0,0
0,1
0,1
0,1
CENAM
0,8
0,2
0,7
0,3
0,2
NRC
-0,9
0,1
0,0
-0,4 -0,1
INMETRO 0,2 -0,1 -0,8 -0,1 -0,2
Bilateral differences between pairs of participants and bilateral normalized errors were calculated according to (2). They are shown in Tables 6A and 6B.
Dij xi x j
Eij 2
Dij
ui2
u
2 j
(2)
Table 6A.1.2 kg standard. Bilateral differences Di j= xi - xj, where xi refers to the result reported by the laboratory in the left column and xj refers to the result reported by the laboratory in the top row.
Dij / mg
LACOMET LATU
INTI
CESMEC CENAM
NRC
LATU
-0,020
INTI
0,336
0,356
CESMEC
-0,040
-0,020
-0,376
CENAM
0,281
0,301
-0,055
0,321
NRC
0,124
0,144
-0,212
0,164
-0,157
INMETRO
0,230
0,250
-0,106
0,270
-0,051
0,106
5/13
Table 6A.2.200 g standard. Bilateral differences Di j= xi - xj, where xi refers to the result reported by the laboratory in the left column and xj refers to the result reported by the laboratory in the top row.
Dij / mg LACOMET LATU
INTI
CESMEC CENAM
NRC
LATU
0,003 0
INTI
0,003 0
0,000 0
CESMEC
0,003 0
0,000 0
0,000 0
CENAM
0,004 9
0,001 9
0,001 9
0,001 9
NRC
0,005 0
0,002 0
0,002 0
0,002 0
0,000 1
INMETRO
0,0000
-0,0030
-0,0030
-0,0030
-0,0049
-0,0050
Table 6A.3.50 g standard. Bilateral differences Di j= xi - xj, where xi refers to the result reported by the laboratory in the left column and xj refers to the result reported by the laboratory in the top row.
Dij / mg LACOMET LATU
INTI
CESMEC CENAM
NRC
LATU
-0,009 0
INTI
-0,003 0 0,006 0
CESMEC
-0,006 0 0,003 0
-0,003 0
CENAM
-0,003 5 0,005 5
-0,000 5
0,002 5
NRC
-0,006 7 0,002 3
-0,003 7
-0,000 7 -0,003 2
INMETRO
-0,010 5 -0,001 5
-0,007 5
-0,004 5 -0,007 0 -0,003 8
Table 6A.4.1 g standard. Bilateral differences Di j= xi - xj, where xi refers to the result reported by the laboratory in the left column and xj refers to the result reported by the laboratory in the top row.
Dij / µg LACOMET LATU
INTI
CESMEC CENAM
NRC
LATU
1,2
INTI
1,1
-0,1
CESMEC
1,1
-0,1
0,0
CENAM
1,0
-0,2
-0,1
-0,1
NRC
0,2
-1,0
-0,9
-0,9
-0,8
INMETRO
0,6
-0,6
-0,5
-0,5
-0,4
0,4
Table 6A.5.200mg standard. Bilateral differences Di j= xi - xj, where xi refers to the result reported by the laboratory in the left column and xj refers to the result reported by the laboratory in the top row.
Dij / µg LACOMET LATU
INTI
CESMEC CENAM
NRC
LATU
0,6
INTI
-0,2
-0,8
CESMEC
0,2
-0,4
0,4
CENAM
0,2
-0,4
0,4
0,0
NRC
-0,1
-0,7
0,1
-0,3
-0,3
INMETRO
0,0
-0,6
0,2
-0,2
-0,2
0,1
Table 6B.1.2 kg standard. Bilateral normalized errors Eij according to (2)
LACOMET LATU
INTI
CESMEC
LATU
0,0
INTI
0,4
0,5
CESMEC
0,0
0,0
-1,0
CENAM
0,3
0,4
-0,2
1,0
NRC
0,1
0,2
-0,9
0,5
INMETRO
0,3
0,3
-0,4
0,7
CENAM
-0,9 -0,2
NRC 0,4
6/13
Table 6B.2.200 g standard. Bilateral normalized errors Eij according to (2)
LACOMET LATU
INTI
CESMEC
LATU
0,1
INTI
0,1
0,0
CESMEC
0,1
0,0
0,0
CENAM
0,2
0,1
0,1
0,1
NRC
0,1
0,1
0,1
0,0
INMETRO
0,0
-0,1
-0,1
-0,1
CENAM
0,0 -0,2
Table 6B.3. 50 g standard. Bilateral normalized errors Eij according to (2)
LACOMET LATU
INTI
CESMEC
LATU
-0,4
INTI
-0,1
0,5
CESMEC
-0,2
0,3
-0,2
CENAM
-0,2
0,7
0,0
0,2
NRC
-0,3
0,1
-0,2
0,0
INMETRO
-0,5
-0,2
-0,6
-0,4
CENAM
-0,2 -0,9
Table 6B.4. 1 g standard. Bilateral normalized errors Eij according to (2)
LACOMET LATU
INTI
CESMEC
LATU
0,3
INTI
0,3
0,0
CESMEC
0,3
0,0
0,0
CENAM
0,3
-0,1
-0,1
0,0
NRC
0,1
-0,3
-0,4
-0,3
INMETRO
0,2
-0,2
-0,2
-0,2
CENAM
-0,4 -0,2
Table 6B.5. 200 mg standard. Bilateral normalized errors Eij according to (2)
LACOMET LATU
INTI
CESMEC CENAM
LATU
0,3
INTI
-0,1
-0,4
CESMEC
0,1
-0,2
0,2
CENAM
0,2
-0,2
0,3
0,0
NRC
0,0
-0,3
0,1
-0,1
-0,2
INMETRO
0,0
-0,4
0,1
-0,1
-0,2
NRC
-0,1 NRC
-0,3 NRC
0,2 NRC
0,1
8. Link to CCM key comparisons In order to demonstrate equivalence, the present comparison was linked to CCM.M.M.K5 [3]. CENAM, INMETRO and NRC have participated in K5 and acted as linking laboratories. Their degrees of equivalence obtained in this comparison (that is, the differencesD0and the corresponding expanded uncertaintiesU0) are shown in table 7. In CCM.M.M.K5 sets of two standards (called JX and JY) were measured for each nominal value. Then, the entries in table 7 correspond to average values of the respective results for JX and JY.
7/13
CENAM INMETRO NRC
Table 7. Degrees of equivalence of the linking laboratories in CCM.M.M-K5
2 kg
200 g
50 g
1g
200 mg
D0 /mg U0 / mg D0 / mg U0 / mg D0 / mg U0 / mg D0 / µg U0 / µg D0 / µg U0 / µg
0,019 0,134 0,000 0,010 0,001 0,006 -0,1 1,7 -0,4 0,8
0,028 1,201 0,003 0,120 0,002 0,030 0,6
1,5
0,2
1,4
-0,084 0,079 0,000 0,009 0,007 0,011 0,7
1,7
0,3
2,0
The reference values of the current comparison, linked to CCM.M.M.K5 (linked reference values LRV) were calculated as:
LRV wC
eC D0,C
wI eI D0,I wN wC wI wN
eN D0,N
where he symbol e refers to mass error, and the subscripts C, I and N refer to CENAM, INMETRO and NRC respectively. The weights wj (for j = C, I or N) were calculated as:
wj
u2
ej
1
Dj ,0
u
2 j
1 U
2 j ,0
4
(3)
For the calculations in (3), the results of the linking laboratories in both comparisons were
considered as non-correlated. The standard uncertainty associated to the linked reference
value is
uLRV
1 wC wI wN
The values of LRV and their associated uncertainties are shown in table 8.
Table 8. Key comparison reference values, linked to CCM.M.MK5
2 kg
200 g
50 g
1g
200 mg
LRV /mg 2,765
0,317
0,0738
0,0034
0,000 1
uLRV /mg
0,054
0,007
0,0036
0,000 6
0,000 4
Then, the degrees of equivalence stated in table 4 were corrected by subtracting the term
NLRV-LRV to the differences D, obtaining linked differences DL. Linked normalized errors
EnL are now calculated as:
EnL 2
DL
u
2 x
u2 LRV
These results are shown in table 9A and 9B. The values of DL and their associated uncertainties are plotted in figures 3A to 3E, where LRV is represented as the zero line and its associated uncertainty as the solid pink lines.
8/13
LACOMET LATU INTI CESMEC CENAM NRC INMETRO
2 kg DL / mg UDL/mg -0,225 0,85 -0,245 0,69 0,111 0,23 -0,265 0,32 0,056 0,17 -0,101 0,15 0,005 0,24
Table 9A. Linked degrees of equivalence
200 g
50 g
1g
DL / mg UDL/mg DL / mg UDL/mg DL / µg UDL / µg
-0,005 0,025 0,006 0,023 -0,2
3,2
-0,002 0,028 -0,003 0,010 1,0
2,9
-0,002 0,023 0,003 0,012 0,9
2,3
-0,002 0,033 0,000 0,012 0,9
3,2
0,000 0,021 0,003 0,009 0,7
1,6
0,000 0,032 0,000 0,016 0,0
1,8
-0,005 0,027 -0,004 0,009 0,4
1,8
200 mg
DL / µg UDL / µg
-0,3
1,5
0,3
1,8
-0,5
1,4
-0,1
2,2
-0,1
1,1
-0,4
1,6
-0,3
1,0
Table 9B. Values of En respect to the linked reference value
2 kg 200 g 50 g 1 g 200 mg
LACOMET -0,3 -0,2 0,3 -0,1 -0,2
LATU
-0,4 -0,1 -0,3
0,3
0,2
INTI
0,5
-0,1
0,3
0,4
-0,4
CESMEC -0,8 -0,1
0,0
0,3
0,0
CENAM
0,3
0,0
0,3
0,5 -0,1
NRC
-0,7
0,0
0,0
0,0
-0,3
INMETRO 0,0 -0,2 -0,5 0,2 -0,3
Figure 1A.Stability of the 2 kg standard
3 e / mg
Figure 1B. Stability of the 200 g standard
0,36
e / mg
2,75
0,32
2,5 July 2009
June 2011
July 2012
Figure 1C. Stability of the 50 g standard
0,085
e / mg
0,075
0,065
July 2009
June 2011
July 2012
0,28 July 2009
June 2011
July 2012
Figure 1D. Stability of the 1 g standard
0,01
e / mg 0,005
0
‐0,005
July 2009
June 2011
July 2012
9/13
Figure 1E. Stability of the 200 mg standard
0,001 e / mg
0
‐0,001
July 2009
June 2011
Figure 2A. Deviations from NLRV for the 2 kg standard. The solid pink lines represent the expanded uncertainty associated to NLRV.
SIM.M.M.K5 2 kg standard
1,2
(e -NLRV) / mg
0,8
0,4
0,0
-0,4
-0,8
-1,2
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
Figure 2B. Deviations from NLRV for the 200 g standard. The solid pink lines represent the expanded uncertainty associated to NLRV.
SIM.M.M.K5 200 g standard
0,04
(e -NLRV) / mg
0,02
0,00
-0,02
-0,04
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
10/13
Figure 2C. Deviations from NLRV for the 50 g standard. The solid pink lines represent the expanded uncertainty associated to NLRV.
SIM.M.M.K5 50 g standard
0,03
(e -NLRV) / mg
0,02
0,01
0,00
-0,01
-0,02
-0,03
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
Figure 2D. Deviations from NLRV for the 1 g standard. The solid pink lines represent the expanded uncertainty associated to NLRV.
SIM.M.M.K5 1 g standard
4
(e -NLRV) / µg
2
0
-2
-4
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
Figure 2E. Deviations from NLRV for the 200mg standard. The solid pink lines represent the expanded uncertainty associated to NLRV.
SIM.M.M.K5 200 mg standard
3
(e -NLRV) / µg
2
1
0
-1
-2
-3
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
11/13
Figure3A. Deviations from LRV for the 2 kg standard. The solid pink lines represent the expanded uncertainty associated to LRV.
SIM.M.M.K5 2 kg standard
1,2
(e -LRV) / mg
0,8
0,4
0,0
-0,4
-0,8
-1,2
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
Figure 3B. Deviations from LRV for the 200 g standard. The solid pink lines represent the expanded uncertainty associated to LRV.
SIM.M.M.K5 200 g standard
0,04
(e -LRV) / mg
0,02
0,00
-0,02
-0,04
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
Figure 3C. Deviations from LRV for the 50 g standard. The solid pink lines represent the expanded uncertainty associated to LRV.
SIM.M.M.K5 50 g standard
0,03
(e -LRV) / mg
0,02
0,01
0,00
-0,01
-0,02
-0,03
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
12/13
Figure 3D. Deviations from LRV for the 1 g standard. The solid pink lines represent the expanded uncertainty associated to LRV.
SIM.M.M.K5 1 g standard
4
(e -LRV) / µg
2
0
-2
-4
LACOMET LATU
INTI CESMEC CENAM NRC INMETRO
Figure 3E. Deviations from LRV for the 200 mg standard. The solid pink lines represent the expanded uncertainty associated to LRV.
SIM.M.M.K5 200 mg standard
3
(e -LRV) / µg
2
1
0
-1
-2
-3
LACOMET LATU
INTI CESMEC CENAM
NRC INMETRO
9. References
1. Picard A., Davis R.S., Gläser M., Fujii K., Revised formula for the density of moist air (CIPM-2007),Metrologia 45 (2008), 149-155
2. Cox, M.G., The evaluation of key comparison data, Metrologia 39 (2002) 589-595 3. Van Andel, I. et al; Report on CIPM key comparison of the second phase of multiples and
submultiples of the kilogram (CCM.M-K5)Metrologia48(2011), Tech. Suppl., 07008
13/13
Ver+/-