ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
1
TRACEABILITY OF RESULTS: DEVELOPMENT OF A SECONDARY CERTIFIED REFERENCE MATERIAL FOR SOMATIC CELL COUNTING FOR MILK QUALITY CONTROL.
Author: Gabriela A. Costamagna*1. Co-authors: Marina S. Toscana1, Lucía M. Marmo1, Javier A. Pugliotti1, Juliana A. Rinaldi1. 1Laboratorio de Materiales de Referencia of the Departamento de Materiales de Referencia, of the Dirección Técnica de Metrología Química of the Instituto Nacional de Tecnología Industrial (INTI), Ruta 34 Km 227- (2300) Rafaela, Santa Fe, Argentina. * Email address: gcostamagna@inti.gob.ar
ABSTRACT
The nutritional value of milk, its quality and acceptance by consumers depend on its composition. Its analysis is essential to support livestock and dairy production, since it provides information on its quality, composition, and safety, and in several countries, determines the payment for quality to the producer. For this reason, laboratories must be technically competent and ensure the validity of their results through metrological traceability to a stable and common international reference, which guarantees the comparability and global acceptance of measurement results. To achieve this, the use of Certified Reference Materials (CRM) is key, since it allows the calibration of rapid measurement equipment, widely used by the industry to obtain results in less time without compromising accuracy. This requires the development of regional capabilities to produce CRM in the dairy sector, contributing to obtaining valid results and improving competitiveness in global markets. In 2020, the Joint Research Center launched the CRM ERM®-BD001, with two levels of Somatic Cell Count (SCC) in spray-dried cow milk in an inert gas atmosphere, which is metrologically traceable to the International System of Units (SI) and ensures global analytical equivalence.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
2
In this framework, the Laboratorio de Materiales de Referencia of INTI-Argentina used ERM®-
BD001 to produce secondary CRMs (INTI-MRC002) in a matrix similar to natural milk, achieving
metrological traceability of the assigned SCC values to the SI through the certified values of the
primary CRM. The INTI-MRC002 met the requirements of ISO 17034:2017 and its associated
guidelines/standards, ensuring the suitability of the production, packaging and storage process,
and ensuring its fitness for use. Recognition of the quality management system as producers of
reference materials was achieved before the QSTF of the SIM. This advancement has strengthened
the metrological traceability chain in Latin America, ensuring the comparability of the results and
their acceptance at an international level.
Keywords: certified reference material (CRM), somatic cell count (SCC), metrological traceability.
INTRODUCTION
Milk is one of the most versatile and valuable foods in the food industry and has been known as the most complete food in nature for millennia (Linehan K. et. al., 2024). According to the Food and Agriculture Organization (FAO), approximately 80% of the annual milk production comes from cows, with the remainder from other dairy animals such as buffalo, goats, camels, and sheep.
The composition of milk determines its nutritional quality, its value as a raw material for the production of food products, and its acceptance by consumers. Its analysis is essential to support livestock and dairy production, as it provides information on its quality, composition, and safety. Furthermore, it enables the identification of factors affecting production, the optimization of livestock feeding and management, and the improvement of production efficiency.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
3
In many countries, programs include the physicochemical, microbiological, and sanitary
analysis of raw milk, enabling the implementation of payment systems based on its quality. These
systems represent one of the approaches used by dairy companies and industries to raise awareness
and incentivize milk producers to improve milk quality. They play a crucial role in determining
the final price paid to producers, which is based on the levels of specific milk components, such
as proteins, somatic cell count, total bacterial count, among others (Revelli, G. R., (2011);
Busanello, M., (2020); Resolución Conjunta N° 739 y 495 de 2011).
For this reason, testing laboratories associated with this activity must be technically competent and ensure the validity of their results through metrological traceability. Traceability to a common and stable international reference guarantees the comparability and global acceptance of measurement results (Murcia-Rubiano, F. et. al.,2021).
Metrological traceability is defined through the International Vocabulary of Metrology (VIM) as property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributes to the measurement uncertainty (Joint Committee for Guides in Metrology [JCGM], 2012).
Comparability of measurements is an essential feature of an international system within measurement results can be universally accepted and can only be guaranteed if those results can be linked to internationally recognized references and comparison through their relationship to that common stable reference (Centro Nacional de Metrología [CENAM], 2002).
Comply whit this, laboratories and industries design and implement different tools of ensuring the validity of its results, among which are, without limiting, participation in proficiency testing and interlaboratory comparisons, internal quality control, analysis of spiked samples and
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
4
use of reference material. The latter is particularly effective for evaluating and demonstrating the
validity of analytical data (Barwick, V. et. al., 2001).
The ISO Guide 30 (2015) defines a reference material (RM) as a material that is sufficiently homogeneous and stable with respect to one or more specified properties, which has been established to be fit for its intended use in a measurement process. Furthermore, defines a Certified reference material (CRM) is a reference material characterized by a metrologically valid procedure for one or more specified properties, accompanied by a reference material certificate that provides the value of the specified property, its associated uncertainty, and a statement of metrological traceability.
Padua-Gandra et. al. (2016) highlights the relevance of method validation processes and the development of internal quality, and traceability of measurements to international systems. Thus, it is necessary to build capabilities within the country for the development and production of CRM that can be used in the food sector. This contribution ensures the validity of results in laboratories and, consequently, enhances competitiveness in the market by achieving valid outcomes that support products in global markets.
The production of RM and CRM is governed by international standards, such as ISO 17034 (2016) and its associated guidelines and is carried out by various institutions, including National Metrology Institutes (NMIs) like the Instituto Nacional de Tecnología Industrial (INTI).
In Argentina, Laboratorio de Materiales de Referencia of INTI-Lácteos headquarters in Rafaela, through its Sistema Centralizado de Calibración (SICECAL®), designs, produces and distributes RM and CRM of different dairy matrices. These materials are acquired and used by national and Latin American laboratories that analyze milk, dairy products and derivatives for calibration, control of their measuring equipment and evaluation of analytical and instrumental INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
5
methods, ensuring the validity of the results they obtain. An example of this is the CRM of somatic
cell count in milk.
The somatic cell count (SCC) in milk is used a crucial diagnostic tool to evaluate udder health status and ensure milk quality and safety. High SCC levels are generally associated with mastitis, an infection of the mammary gland that impacts milk quality and profitability, leading to losses due to reduced production, lower quality and safety, and increased veterinary costs. (Schukken et al., 2003; International Dairy Federation [IDF], 2021, p.15; Baltián, L. R. et al., 2023).
The importance of raw milk quality is undeniable in dairy production, since it is closely linked to flavor, sensory and physicochemical aspects, as well as hygiene, nutrition and safety of the final dairy products (Ntuli et. al., 2023).
Different methods are used to determine SCC in milk. There are "direct" methods, which involve counting particles or cells, and "indirect" methods, which estimate the number of cells based on reactions revealing cellular components. The direct microscopic counting is considered the reference method by the International Organization for Standardization (ISO) y la International Dairy Federation (IDF) according to ISO 13366-1:2008/Cor 1: 2009 | IDF 148-1:2008/Cor 1: 2009. This method is laborious and requires a very well-trained analyst. In contrast, among the indirect methods, one of the automated SCC methods are based on flow cytometry, which stains cells and detects the stained particles using light energy emitted by them. In these methods, a sample is mixed with a buffer and a dye, then transferred to a rotating disk illuminated by a Zenon lamp. The light emitted by the cells is filtered and detected by a photomultiplier, generating an electrical pulse that is amplified (Di Marzo, L., 2016). According to the provisions of the reference standard ISO 13366-2:2006 | IDF 148-2:2006, periodic calibration of these devices is essential to guarantee
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
6
their precision. This method is widely used by dairy laboratories, allowing large-scale counts at
low cost and with considerably greater precision. However, said reference standard raises the issue
of the lack of availability of a CRM that is an integral part of a reference system for SCC in milk.
Since 2008, la IDF, International Committee for Animal Recording (ICAR) and European Commission’s Joint Research Centre (EC JRC) have closely collaborated in a joint project to develop solutions and tools to promote a better global equivalence in SCC in milk.
Thus, in 2020, the JRC launched an CRM (CRM ERM®-BD001) with 2 levels of SCC in milk powder in an inert gas atmosphere. As SCC is a method-dependent measurand, the characterization of this CRM was conducted through a network of laboratories that applied the same reference measurement procedure (ISO 13366-1:2008/Cor 1: 2009 | IDF 148-1:2008/Cor 1: 2009).
Furthermore, the producer of the CRM declares that the certified value is traceable to SI units, which is based on the traceability of all relevant input factors, such as the calibrated micropipette used for milk sampling or the slide pre-marked with an outline shape (circular or rectangular) with a defined area. Since the assigned values are a combination of consistent results individually traceable to the SI, the assigned quantity values themselves are also traceable to the SI. The use of CRM allows a direct connection to be established to the SI, ensuring analytical equivalence on a global scale and providing an anchor point to which secondary reference materials can be linked (EC Joint Research Centre, 2020)1.
In this context, the Laboratorio de Materiales de Referencia of INTI-Argentina used ERM®-BD001 to produce secondary CRMs (INTI-MRC002) in a matrix similar to natural milk,
1 Metrological traceability of the primary reference material ERM®-BD001 declared in the certification report by the producer. Doi:10.2760/681742
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
7
achieving the metrological traceability of the assigned SCC values to the SI through the certified
values of the primary CRM. The expectation for this was to promote better equivalence of SCC
results following the metrological traceability chain.
MATERIALS AND METHODS
Selection of raw material
To achive 5 candidate CRM with low, medium and high SCC values covering a range between 50 000 cells/ml and 1 200 000 cells/ml established by the reference standard, raw milk from authorized dairy farms and industries was acquired, and selected once the SCC and its suitability for use were determined.
Analytical Method
The analytical methods utilized for assigning values adhere to the ISO-IDF Standards.
ISO 13366-1:2008/Cor 1: 2009 | IDF 148-1:2008/Cor 1: 2009. Milk ˗˗ Enumeration of somatic cells ˗˗ Part 1: Microscopic method (Reference method). The Laboratory is accredited by the Organismo Argentino de Acreditación (OAA) for its application of this method.
ISO 13366-2:2006 | IDF 148-2:2006. Milk ˗˗ Enumeration of somatic cells ˗˗ Part 2: Guidance on the operation of fluoro-opto-electronic counters. These measurements are conducted in external subcontracted laboratories, accredited by the OAA.
CRM Preparation
After selecting the raw material, the necessary mixtures were prepared to create 5 sublots each falling within the following ranges: 50 000 cells/ml and 200 000 cells/ml, 201 000 cells/ml
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
8
and 400 000 cells/ml, 401 000 cells/ml and 600 000 cells/ml, 601 000 cells/ml and 800 000 cells/ml
and 801 000 cells/ml and 1 200 000 cells/ml. These sublots were then analyzed using the BacSomaticTM FOSS analyzer, which conforms to the method described in ISO 13366-2:2006 |
IDF 148-2:2006. Each sublot was labeled as INTI-MRC002-G2304 from 1 to 5.
Each subbatch of candidate CRM was filtered to remove any suspended “macroparticles”. Subsequently, it was transferred to glass jars and subjected to heat treatment using a flowing steam autoclave to eliminate bacterial flora.
The conditioned candidate CRM was homogenized in a thermostated water bath with gentle stirring by inversion. Working in an aseptic environment, 2-bromo-2-nitro-1,3-propanediol (Bronopol) was added as a preservative and the mixture was divided into sterile 50 ml plastic bottles, each labeled accordingly.
Homogeneity Study
Following the recommendations of ISO Guide 35 (2017), a homogeneity study was conducted, involving the analysis from each sublot, a subset of 5 units, employing a simple random sampling strategy. The determination of SCC was then performed on these units in a single measurement run, in a completely random order, using the BacSomatic™ FOSS analyzer.
The analytical results obtained were evaluated using the Grübbs test for outlier detection. Trend analysis of the results was conducted regarding the packaging order and the order of analysis execution, followed by the assessment of intra-unit and inter-unit homogeneity through one-way analysis of variance (ANOVA), applying F-test at a significance level of 0,05.
The homogeneity uncertainties for each sublot were estimated following the guidelines of ISO Guide 35 (2017), applying the formulas (1), (2), (3), and (4):
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
9
ℎ = = ,
(1)
=
(2)
2
=
−
(3)
,2
=
√2
(4)
Between-unit component of variance from a homogeneity study, expressed as a
standard deviation.
Within-unit mean squares degrees of freedom. Number of observations per unit.
Mean square or variance between units.
Mean square or variance within the unit.
, Standard uncertainty associated with between-unit variability.
Stability Study
A long-term stability study was conducted, performing 2 measurements at each time point (1, 8, 15, 22, 29, 36, 43, 50, 57, and 62 days), under storage conditions of 4 °C ± 2 °C. For the statistical analysis of the results, a linear regression analysis of variance was conducted, considering the case of linear behavior and applying Fisher's F-test at a significance level of 0,05.
Stability uncertainties were estimated for each sublot following the guidelines of ISO Guide 35 (2017), applying formula (5).
= 1 × ( 1 + )
(5)
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
10
1 Standard deviation associated with the slope β1 of the regression (standard error for the estimated slope)
1 Time interval between value assignment and the initial stability monitoring point in days.
Period of validity of certificate issued.
Characterization
To conduct the analytical measurement, individual samples of CRM candidates are randomly selected.
Characterization was performed using the reference measurement procedure at the Laboratorio de Materiales de Referencia and the fluoro-opto-electronic method was employed in multiple accredited laboratories.
Characterizations uncertainty was estimated by the weighed the standard uncertainty of the reference method results ( ) with the standard uncertainty of the fluor-opto-electronic method results ( ), following formula (6).
ℎ = 0,5 × + 0,5 ×
(6)
Primary Reference Material Used for Trueness Assessment
The CRM ERM®-BD001, traceable to the International System of Units (SI), was used as the primary reference material to evaluate the trueness of the reference measurement procedure. For this purpose, it was prepared and analyzed by reference, according to the supplier's instructions.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
11
Trueness was evaluated by comparing the result obtained by the laboratory with the value
assigned to the CRM ERM®-BD001 declared in its certificate, using the formula (7) as the
acceptance criterion.
| − | ≤ √ 2 + 2
(7)
Average value obtained in the CRM ERM®-BD001 measurement. SCC certified value. Standard uncertainty associated with the average value obtained by measuring the
SCC of the CRM ERM®-BD001.
Standard uncertainty associated with the certified value of SCC.
Value Assignment and Uncertainty Estimation
Each assigned value was calculated by weighed the averages of the results obtained according to both the reference method ( ) and the routine method ( ), after evaluating the presence of outliers according to formula (8).
= 0,5 × + 0,5 × (8)
The uncertainty of the assigned value was estimated by a quadratic combination of the standard uncertainty from the characterization ( ℎ ), the standard uncertainty provided by the homogeneity study ( ℎ ) and the standard uncertainty provided by the long-term stability study ( ). A coverage factor k = 2 (95% confidence) was applied to obtain the expanded uncertainties, as indicated by the following formula (9).
= 2 √( ℎ )2 + ( ℎ )2 + ( )2
(9)
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
12
RESULTS AND DISCUSSION
Homogeneity Study
The results for the homogeneity study were subjected to the Grübbs test, to detect outlier values in the dataset. In the analyzed batch, no outlier values were detected, which allowed the initial dataset to be used for subsequent statistical analysis
The absence of trends in the packaging and in the measurement sequence was verified through a linear regression study, where it was established as an acceptance criterion that the confidence interval of the slope of the data included the value zero. Table 1 presents the results of the trend analysis for each sublot (1 to 5), where the absence of significant trends can be observed both during the packaging of the production batch and in the material analysis sequence.
Finally, the homogeneity of the data was verified through one-way ANOVA. Table 1 presents the results of the analysis for each of the sublots. In all cases, the FCalculated value was lower than the FCritical value, allowing us to confirm that the means of the SCC values of the 5 analyzed containers are equal. This confirms the homogeneity of the SCC parameter in the material.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
13
Table 1.
Results of the trend analysis and homogeneity for the 5 sublots.
ANALYZED TREND
TEST ANOVA
Measurement trends (analyzed)
Processing trends (packaging)
UNCERTAINTY
slope interval
Lower
Upper
1 -668,6
479,8
2 -1762,0
1531,3
3 -2266,4
1476,1
4 -2306,6 5 -4765,0
2474,4 2030,8
No significant trends
were observed
slope interval
Lower
Upper
-712,5
425,8
-1574,4
1721,3
-1139,7
2517,3
-3092,4 -4206,0
1567,9 2786,4
No significant trends
were observed
FCalculated
FCritical
0,330 0,124 0,142 1,397 1,017
4,387 4,387 4,387 4,387 4,387
FCalculated < FCritical
uhom
(cells/ml)
1 906 5 843 6 659 5 018 1 673
Stability Study
Following the guidelines of the reference guide for data evaluation, a regression analysis of SCC in contrast to time and a one-way analysis of variance were conducted, with the acceptance criterion that the FCalculated value should be lower than the critical value of F(1,n-2,α). Table 2 presents the results of the variance study, while Figure 1 illustrates the absence of trends during the evaluated period and conditions. As a result, no significant degradation was observed, demonstrating the material's stability under these conditions. Additionally, the uncertainties associated with long-term stability are depicted.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
14
Table 2.
Data obtained from the analysis of variance for the study of stability.
SAMPLE
1 2 3 4 5
REGRESSION DATA
FCalculated
F(1,n-2, α)
1,546
5,320
1,703
5,320
0,850
5,320
1,597
5,320
2,289
6,610
ults (cells/ml)
1 796 4 022 7 376 8 472 45 925
Figure 1. Long-term stability study.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
15
Trueness Assessment
Table 3 presents the results obtained by the laboratory during the analysis of the CRM ERM®-BD001, including the assigned value reported in the certificate and the assessment of its precision.
Measurements and property values are demonstrated to be mutually consistent within their respective uncertainties.
Table 3.
Trueness Assessment.
Reference Value ± U*
64 000 ± 8 000
1 202 000 ± 121 000
Laboratory Value ± U*
75 000 ± 31 228 58 000 ± 31 228
1 135 000 ± 31 228
(cells/ml)
Bias
11 000 -6 000 -67 000
1 112 000 ± 31 228 -90 000
Note. *Expanded uncertainty
Reproducibility Limit 35 160 27 261 441 967
433 126
11 000 ≤ 32 236 6 000 ≤ 32 236 67 000 ≤ 14 965
90 000 ≤ 124 965
Material Value Assignment and CRM Uncertainty Estimation
Table 4 presents the expanded uncertainties of the CRMs estimated by combining the uncertainties from the studies mentioned above and using a coverage factor (k = 2) with a significance level of 95 %.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
Table 4.
CRM Uncertainty Estimation.
Sample
1 2 3 4 5
Characterization Uncertainty 15 125 32 107 35 931 42 183 64 378
(cells/ml)
Homogeneity
Stability
Uncertainty
Uncertainty
1 906
1 795
5 843
4 022
6 659
7 376
5 018
8 472
1 673
19 374
Combined Uncertainty 15 349 32 881 37 280 43 317 67 251
16
Expanded Uncertainty 30 698 65 762 74 560 86 634 134 502
Finally, the reference value of the CRM in SCC assigned through the characterization study was 104 056 cells/ml ± 30 698 cells/ml, 356 041 cells/ml ± 65 762 cells/ml, 493 864 cells/ml ± 74 560 cells/ml, 711 346 cells/ml ± 86 634 cells/ml y 905 160 cells/ml ± 134 502 cells/ml for CRM 1, 2, 3, 4 y 5 respectively, completing the pack provided by the Laboratorio de Materiales de Referencia.
Figure 2: CRM of SCC produced by the Laboratorio de Materiales de Referencia, INTI.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
17
CONCLUSIONS
Based on the results, the Laboratorio de Materiales de Referencias of INTI has achieved the development of a secondary CRM in SCC, achieving metrological traceability of the values assigned to the SI through the certified values of the primary CRM ERM®-BD001.
The laboratory followed the guidelines set by ISO 17034:2017 and its associated standards/guidelines, which allowed the developed CRM INTI-MRC002-G2304 to meet the requirements for homogeneity and long-term stability, thus confirming the suitability of the production, packaging, and storage processes.
By demonstrating the trueness of the microscopic method, it was confirmed that the results obtained using this method were metrologically traceable to the same reference as the property values of the primary CRM, in this case to the SI.
The Quality Management System supporting measurement services: Somatic Cell Count, milk (Certified Reference Materials), as defined in a motion of the Quality System Task Force (QSTF), has been reviewed and found to be in conformance with the requirements of ISO 17034:2016 under the mandate of the Inter-American Metrology System in support of the expectations of the CIPM Mutual Recognition Arrangement.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
18
This development has enabled reaching a higher common link in the metrological
traceability chain in Latin America, allowing comparability of results and achieving universal
acceptance. The Laboratorio de Materiales de Referencia, by producing repeated batches, will
ensure the continuous availability of these secondary CRMs for use. Thus, SICECAL® user
laboratories employing our secondary CRM for the calibration of their fluoro-opto-electronic
equipment will obtain measurement results traceable to the SI, which can be compared with results
from other laboratories, enhancing acceptance and mutual trust in the dairy sector.
The use of INTI-MRC002 represents a key tool for ensuring reliable and metrologically traceable measurements. This allows producers to meet industry demands and access fairer economic compensation for their production. This contributes to the sustainable development of dairy farms and the improvement of the quality of milk used for product manufacturing, boosting the competitiveness of the sector.
REFERENCES
Baltián, L. R., Remirez, P., Peratta, D., Schmidt, E. E., Palezza, J., Patrilla, J., Lema Vincens, M., & Portada, J. (2023). Análisis de asociación de alelos del gen BoLA-DRB3.2 con rasgos de producción lechera y número de células somáticas en ganado Holstein de La Pampa. Ciencia Veterinaria, 25(2), 137–150. Doi:10.19137/cienvet202325203.
Barwick, V., Burke, S., Lawn, R., Roper, P., & Walker, R. (2001). Applocations of reference materials in analytical chemistry. Cambridge: Laboratory of the Govemment Chemist.
Busanello, M., Pereira Haygert-Velho, I. M.., Piuco, M. A., Heck, V. I., Stürmer, M., Cosmam, L. C., & Velho, J. P. (2020). Relationship between seasonal variation in the composition of bulk tank milk and payment based on milk quality. Slovak J. Anim. Sci., 53, 2020 (3): 132– 144.
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
19
Centro Nacional de Metrología (CENAM) (2002). Trazabilidad en las mediciones químicas.
Dirección de Metrología de Materiales. Centro Nacional de Metrología, México D.F. Di Marzo, L., Wojciechowski, K. L., & Barbano, D. M. (2016). Preparation and stability of milk
somatic cell reference materials. Journal of Dairy Science. Doi:10.3168/jds.2016-10848. EC Joint Research Centre (2020). Certification report. The certification of the concentration of
somatic cells (somatic cell count, SCC) in cow’s milk: ERM®-BD001. European
Commission. Available throught: https://crm.jrc.ec.europa.eu/p/ERM-BD001 (Visited January 2025). International Dairy Federation (2021). IDF Annual Report 2020-2021: Achieving High Standards (p. 15). International Organization for Standardization (ISO) (2008). Milk-Enumeration of somatic cellsPart 1: Microscopic method (Reference method) (ISO Standard No. 13366-1: 2008/Cor
1:2009 | IDF 148-1:2008/Cor 1:2009) ISO (2006). Milk-Enumeration of somatic cells-Part 2: Guidance on the operation of fluoro-opto-
electronic counters. (ISO Standard No. 13366-2:2006 | IDF 148-2:2006) ISO (2015). Reference materials – Selected terms and definitions (ISO Guide 30:2015). ISO (2016). General requirements for the competence of reference material producers (ISO
Standard No.17034:2016). ISO (2017). Reference materials – Guidance for characterization and assessment of homogeneity
and stability (ISO Guide 35:2017). ISO/IEC (2017). General Requirements for the competence of testing and calibration laboratories
(ISO/IEC Standard No. 17025:2017). Joint Committee for Guides in Metrology (2012). JCGM 200:2012. International vocabulary of
metrology – Basic and general concepts and associated terms (VIM) (3rd ed.). Joint
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
20
Committee for Guides in Metrology. https://www.bipm.org/en/publications/guides/jcgm-
200.
Linehan, K., Patangia, D. V., Reynolds, P. R., & Stanton, C. (2024). Production, composition and
nutritional properties of organic milk: A critical review. Foods, 13(4), 550.
Doi:10.3390/foods13040550.
Murcia-Rubiano, F., Ávila-Guzmán, O., Priero-Navarreta, L., & Rodríguez-Sandoval, P. (2021).
Desarrollo y evaluación de un material de referencia de leche en polvo entera. Informador
Técnico, 86(1), 18 - 33. Doi:10.23850/22565035.3775.
Ntuli, V., Sibanda, T., Elegbeleye, J.A., Mugadza, D.T., Seifu, E., & Buys, E.M. (2003). Dairy
production: Microbial safety of raw milk and processed milk products (chapter 30) Present
Knowledge in Food Safety (pp. 439-454). Elsevier.
Padua-Gandra, R. L., Dos Santos, A. L. S., Dos Santos, P. P. B., Alvarenga, R., Gonçalves-
Junqueira, R., Labanca, R., & Carvalho de Souza, S. (2016). Production and evaluation of
a reference material for moisture, ash, and total fat mass fractions, and titratable acidity
in whole milk powder. Accreditation and Quality Assurance, 21 (1), 47-55.
Doi:10.1007/s00769-015-1179-3.
Resolución Conjunta 739 y 495 de 2011 [Ministerio de Agricultura, Ganadería y Pesca y
Ministerio de Economía y Finanzas Públicas]. Por la cual se establece la creación del
Sistema de Pago de la Leche Cruda sobre la base de Atributos de Calidad Composicional
e Higiénico-Sanitarios en Sistema de Liquidación Única, Mensual, Obligatoria y
Universal. 10 de Agosto de 2011.
https://servicios.infoleg.gob.ar/infolegInternet/anexos/185000-189999/185939/texact.htm
(Visited January 2025).
INTI – Chemical Metrology
ICAR ANNUAL CONFERENCE AND IDF/ISO ANALYTICAL WEEK 2025
21
Revelli, G.R., Sbodio, O. A., & Tercero, E. J. (2011). Estudio y evolución de la calidad de leche
cruda en tambos de la zona noroeste de Santa Fe y sur de Santiago del Estero, Argentina
(1993–2009). Revista de Investigaciones Agropecuarias RIA, (37), 128-139.
Schukken, Y. D., Wilson, F., Welcome, L., Garrison-Tikofsky, & Gonzalez. R. (2003). Monitoring
udder health and milk quality using somatic cell counts. Vet. Res. 34:579–596.
INTI – Chemical Metrology
Ver+/-